Digital Breast Tomosynthesis

Digital breast tomosynthesis is a three dimensional form of mammography. The X-ray tube moves in an arc over the compressed breast, and many low dose projection images are taken. These are rebuilt into thin slices, usually 1 mm thick, that are scrolled through on the monitor.

Overlapping tissue is separated, so a cancer is less likely to be hidden and a normal overlap is less likely to look like a lesion.

The arc is 15 to 50 degrees, with 9 to 25 projections, in 4 to 10 seconds. The usual mammography views are taken (craniocaudal and mediolateral oblique).

Many systems also produce a synthetic two dimensional image from the data, which removes the need for a separate mammogram. The mean glandular dose is about 1 to 2 mGy for each view, and similar to digital mammography when synthetic images are used.

When a full mammogram is also taken, the dose is nearly doubled.

In screening studies, tomosynthesis finds 20 to 40% more invasive cancers, and lowers recalls by 15 to 30%. The gain is largest in women with dense breasts and in women under 60.

Sensitivity is about 85 to 90% and specificity about 90 to 95%.

Indications.

  • Screening of asymptomatic women, in place of or alongside digital mammography.
  • Recall after an abnormal screening mammogram, to confirm or dismiss an overlap of tissue.
  • Women with dense breasts.
  • Symptomatic women aged 40 and over, as a diagnostic test.
  • Characterisation of a mass, asymmetry or distortion.
  • Localisation of a lesion for biopsy (tomosynthesis directed biopsy) and for surgery.
  • Women with breast implants (additional views with the implant displaced).
  • Follow up after breast cancer treatment.

Contraindications and precautions.

  • Pregnancy: avoided for screening, with ultrasound used first.
  • Women under 30: rarely used.
  • The same cautions for compression, implants, recent surgery, and deodorant or powder apply, as they do for digital mammography.
  • Metal clips, wires and markers produce streak artefacts in the slices. Their position is noted.
  • Longer compression time (up to 10 seconds) may be uncomfortable.
  • Total radiation dose should be kept low by using synthetic images instead of a separate mammogram.
  • Reading takes longer, and trained readers are needed.
  • The data storage and display equipment are costly.

Preparation and technique.

Preparation is identical to that for mammography. The breast is positioned and compressed, and the tube moves while projections are taken.

The reconstructed slices are displayed in a stack that is scrolled, along with the synthetic two dimensional image. A lesion is checked in at least two views, and its size and position are noted for biopsy.

Quality checks cover positioning, motion, and the completeness of coverage.

Comparison is made with earlier two dimensional and three dimensional studies. Reading is by one or two radiologists, with software support available in some units.