Interpretation.
- Resting energy expenditure is compared with the predicted value and with the energy delivered.
- A respiratory quotient near 0.7 indicates fat oxidation, near 0.85 mixed fuel use, and near 1.0 predominant carbohydrate oxidation.
- A respiratory quotient over 1.0 suggests overfeeding with lipogenesis. A value under 0.7 usually reflects a measurement error or a metabolic disturbance such as ketosis.
Causes of high energy expenditure. Sepsis, burns, trauma, fever, hyperthyroidism, and agitation.
Causes of low energy expenditure. Sedation, hypothermia, hypothyroidism, starvation, and neuromuscular blockade.
Critical values and findings needing immediate action. A respiratory quotient over 1.0 with raised carbon dioxide production in a patient with respiratory failure requires prompt reduction of carbohydrate load and review of ventilation.
Limitations and artefacts. The test measures a short period, which may not represent a full day. Equipment requires calibration, and leaks distort values.
Follow up tests. Serum prealbumin, C-reactive protein, urea and nitrogen balance, glucose, triglycerides, and blood gases.
Special groups. Children and neonates need devices suited to low flows. Patients with obesity are measured against actual weight, not adjusted weight.