Back

Transcranial Doppler

Transcranial Doppler (TCD) is a non-invasive, bedside pulsed wave Doppler ultrasound examination. It measures blood flow velocity in the large intracranial arteries and the extracranial internal carotid, through natural acoustic windows in the skull. A low frequency (1 to 2 MHz) phased array probe is used.

  • The transtemporal window (over the zygomatic arch) shows the middle cerebral artery (MCA, depth 45 to 65 mm), the anterior cerebral artery (ACA, 60 to 75 mm), the posterior cerebral artery (PCA, 60 to 70 mm) and the terminal internal carotid artery.
  • The transorbital window shows the ophthalmic artery and carotid siphon (depth 40 to 60 mm).
  • The transforaminal (suboccipital) window shows the vertebral arteries and basilar artery (depth 60 to 100 mm and beyond).

Transcranial colour coded duplex (TCCD) combines B-mode imaging of intracranial structures with colour Doppler. Measurements are mean, peak systolic and end diastolic velocity, the pulsatility index (PI), and the direction of flow. Emboli can be detected as high intensity transient signals, and a bubble study (agitated saline) can detect right to left shunting.

The examination takes 20 to 40 minutes at the bedside, with no radiation.

TCD was introduced for monitoring vasospasm after subarachnoid haemorrhage, and for stroke screening in children with sickle cell disease. Cerebral vasospasm occurs in up to 70% of angiographic cases of subarachnoid haemorrhage (symptomatic in about 30%), usually from day 3 to day 14.

In children with sickle cell anaemia, an MCA time averaged maximum velocity of 200 cm each second or more is associated with a stroke risk of around 10% each year without transfusion, and transfusion therapy reduces the risk by over 90%. TCD is operator dependent, but is repeatable, cheap and available at the bedside.

Indications. Vasospasm after subarachnoid haemorrhage (daily monitoring from day 3 to 14). Stroke risk screening in children with sickle cell disease aged 2 to 16 years. Detection and grading of intracranial arterial stenosis or occlusion in stroke and transient ischaemic attack, and monitoring of recanalisation after thrombolysis.

Detection of microembolic signals (carotid disease, cardiac sources, after procedures). Detection of right to left shunt (patent foramen ovale) with agitated saline. Assessment of cerebral autoregulation and vasomotor reactivity. Support in the diagnosis of brain death (absent diastolic flow or reverse flow, systolic spikes). Monitoring during carotid endarterectomy and cardiac surgery.

Raised intracranial pressure (high PI). Traumatic brain injury (vasospasm, hyperaemia, low flow). Cerebral venous thrombosis (limited). Meningitis (vasculopathy). Arteriovenous malformations (feeder vessels).

Contraindications and precautions. None. The temporal window is absent in 5 to 15% of people (higher in older adults, women and people of African or Asian descent). Pressure on the eye is avoided with the orbital window, using low power output (under 10% of the maximum) and short exposure. Head trauma with open wounds, scalp dressings and craniectomy need modified technique.

Normal values vary with age, sex, haematocrit, PaCO2, blood pressure and the cardiac output, and these are recorded.

Preparation and technique. The patient lies supine or sits. The probe is placed over the temporal window (anterior, middle and posterior windows), and the MCA is found at 50 to 55 mm depth with flow towards the probe. Flow direction is described by depth: MCA and ACA flow towards and away from the probe respectively at the bifurcation (about 60 to 65 mm).

The vessel is followed along its depth and the highest velocity is recorded, with angle correction minimal (TCD assumes a near parallel angle). The Lindegaard ratio (MCA mean velocity divided by the extracranial ICA mean velocity measured in the neck) is used to distinguish hyperaemia from vasospasm. Microembolic detection is done with a headset for 30 to 60 minutes.

A bubble study involves injecting agitated saline during Valsalva release while monitoring MCA for bubble signals.

Related

Clinical toolsCalculators

Latest content

Atlas’ Videos

On the evidence

Doses are traceable.

Management sections cite the guideline they came from. Atlas supports clinical judgement rather than replacing it; verify against current national guidance and the patient in front of you.

Learn it.
Know it.
Own it.