Carboxyhaemoglobin and Methaemoglobin

Carboxyhaemoglobin (COHb) and methaemoglobin (MetHb) are dyshaemoglobins measured by multiwavelength co-oximetry on arterial or venous whole blood, usually on the blood gas analyser, in 1 to 2 minutes. Co-oximetry measures absorbance at four or more wavelengths and reports oxyhaemoglobin, deoxyhaemoglobin, COHb and MetHb as a percentage of total haemoglobin.

Pulse co-oximeters estimate COHb (SpCO) and MetHb (SpMet) non-invasively but are less accurate (error of plus or minus 3 to 5 percentage points). Conventional two wavelength pulse oximeters cannot distinguish these species: they read COHb as oxyhaemoglobin (falsely normal saturation) and MetHb as a fixed reading near 85%.

Carbon monoxide binds haemoglobin with about 240 times the affinity of oxygen, shifts the dissociation curve to the left and inhibits cytochrome oxidase, causing tissue hypoxia despite normal oxygen tension.

It is the commonest cause of fatal poisoning worldwide (fires, faulty heaters, charcoal braziers, generators, car exhaust, indoor cooking with biomass fuel), with about 1 to 2 deaths in each 100 000 people in many countries, and delayed neurological sequelae in up to 40% of survivors of serious poisoning.

Methaemoglobinaemia results from oxidation of iron to the ferric state, either congenital (cytochrome b5 reductase deficiency, haemoglobin M) or acquired (drugs and chemicals: dapsone, local anaesthetics such as benzocaine and prilocaine, nitrites, nitrates in well water and food, aniline dyes, primaquine, rasburicase, sulphonamides, nitric oxide), and causes cyanosis that does not improve with oxygen.

Indications. Carbon monoxide: exposure to fire, smoke or enclosed space combustion; headache, dizziness, nausea, confusion, syncope, chest pain, or several household members unwell together; coma of unknown cause; pregnancy with exposure.

Methaemoglobin: cyanosis with normal oxygen tension, chocolate brown blood, oxygen saturation of 85% that does not respond to oxygen, after exposure to drugs or chemicals, infants with diarrhoea or exposure to nitrate rich water, G6PD deficiency with haemolysis. Haemoglobin variants as the cause of low pulse oximeter readings.

Not for chronic low level exposure screening, which uses breath carbon monoxide.

Specimen. Arterial or venous heparinised whole blood 0.5 to 1 mL in a syringe, air excluded, analysed immediately (COHb is stable for hours at 4 °C in a sealed tube, but MetHb rises on storage and should be measured within 30 minutes).

Record the time of exposure, the time of sampling and any oxygen therapy, because high flow oxygen shortens the half life of COHb from 4 to 6 hours in room air to about 60 to 90 minutes. A sample after prolonged oxygen therapy underestimates the exposure.