Acoustic Reflex Testing

The acoustic reflex is the contraction of the stapedius muscle in the middle ear in response to a loud sound. The contraction stiffens the ossicular chain and reduces the sound reaching the inner ear, which protects it to some degree from loud noise.

Acoustic reflex testing measures this reflex. It tests the middle ear, the cochlea, the auditory nerve, the lower brainstem, and the facial nerve in one test.

The reflex arc starts in the cochlea, passes along the auditory nerve to the cochlear nucleus in the brainstem, crosses to the facial nerve nuclei on both sides, and returns along the facial nerve (which supplies the stapedius muscle) to the middle ear.

A loud tone is presented to one ear. The stiffening of the eardrum caused by stapedius contraction is detected by the same probe used for tympanometry, as a small decrease in compliance.

The sound is presented to the same ear as the probe (ipsilateral) and to the opposite ear (contralateral). Frequencies are 500, 1000, 2000 and 4000 Hz, and broadband noise.

The test takes 5 to 10 minutes, and follows tympanometry, since the middle ear must be normal for a reliable test. In normal hearing, the reflex threshold is about 70 to 100 dB HL, which is about 85 dB above the hearing threshold. Reflex decay testing measures whether the reflex is sustained, by presenting a tone at 10 dB above the threshold for 10 seconds, at 500 and 1000 Hz.

Indications.

  • Differentiating the site of the lesion in hearing loss: middle ear, cochlea, auditory nerve, or brainstem.
  • Suspected retrocochlear lesion (vestibular schwannoma and other cerebellopontine angle lesions): reflex decay and absent reflexes.
  • Facial nerve palsy, to locate the lesion in relation to the branch to the stapedius muscle and to give an idea of the outlook.
  • Confirmation of the middle ear status, together with tympanometry, in conductive hearing loss and otosclerosis.
  • Estimation of hearing in children and in patients who cannot cooperate with behavioural tests (acoustic reflex threshold is used as a cross-check).
  • Recruitment (loudness growth) assessment in cochlear loss, and detection of non-organic hearing loss, when reflex thresholds are better than the reported hearing thresholds.
  • Brainstem and neurological disease (multiple sclerosis, brainstem stroke, tumours), where reflexes help to locate lesions.
  • Myasthenia gravis (fatigable reflex), and other neuromuscular disorders in selected cases. Not used alone for diagnosis, and not used for screening in infants under about 4 to 6 months, in whom the reflex may be unreliable.

Contraindications and precautions.

  • Loud sounds up to 100 to 110 dB HL are presented, so the test is avoided or limited in patients with hyperacusis, painful loudness intolerance, or very low discomfort levels, and in those with noise-induced tinnitus that is aggravated by loud sounds.
  • Recent ear surgery (stapes surgery), and ear discharge, perforations, or blocked ear canal: no valid test.
  • Patients who cannot keep still or tolerate the probe.
  • Very recent ear infection or acute otitis media: the test is delayed.
  • A normal tympanogram (type A) is needed. A middle ear abnormality that prevents the probe from detecting the contraction (flat curve) makes the reflex test invalid.
  • Test time and sound levels are kept to the minimum necessary to avoid damage, and equipment is calibrated.
  • Neonates and young infants have a high-frequency probe tone with careful interpretation.

Preparation and technique.

Otoscopy and tympanometry are done first. The probe is sealed in the ear canal, and the pressure is set to the peak of the tympanogram, where the eardrum is most mobile.

Tones are presented at 500, 1000, 2000 and 4000 Hz, starting at about 70 to 80 dB HL, and raised in 5 dB steps. The response is a small, repeatable decrease in compliance (a deflection of at least 0.02 mL), time-locked to the sound.

The lowest level at which the response occurs is the reflex threshold. Ipsilateral and contralateral reflexes are measured for each ear. For reflex decay, a tone at 500 and 1000 Hz at 10 dB above the threshold is presented for 10 seconds, and the amplitude is observed to check whether it falls to half or less. The test is stopped at 110 dB HL if there is no response, or if the patient is uncomfortable.