Otoacoustic Emissions
Otoacoustic emissions are faint sounds produced by the outer hair cells of the cochlea, either spontaneously or in response to a sound. The test is quick, objective, and needs no response from the patient.
Outer hair cells amplify sound inside the cochlea, and the amplified energy travels back through the middle ear to the ear canal, where a sensitive microphone can record it. A normal emission means that the middle ear and the outer hair cells of the cochlea work.
A small probe with a loudspeaker and a microphone is placed in the ear canal. Two types of emission are recorded. Transient evoked emissions (TEOAE) are produced in response to a brief click, are recorded over about 20 milliseconds, and cover about 500 to 4000 Hz.
Distortion product emissions (DPOAE) are produced in response to two pure tones of different frequencies. The response arises at a frequency determined by the two tones, and can be tested at specific frequencies from about 1000 to 8000 Hz, giving a frequency-specific result.
Results are shown as a pass (emission present, with enough signal over the background noise) or refer, or as a plot of the amplitude by frequency (DP-gram).
The test takes 1 to 5 minutes for each ear, and is done in a quiet room, with a sleeping or quiet baby, or with an adult sitting still.
Otoacoustic emissions are used for newborn hearing screening throughout the world, since permanent hearing loss affects about 1 to 3 in 1000 babies. Emissions are present in nearly all ears with normal hearing, and are absent when hearing loss is over about 30 to 40 dB HL, with a sensitivity of about 90 to 95% for permanent hearing loss.
Indications.
- Newborn and infant hearing screening, in the first days or weeks of life.
- Follow-up screening of children at risk of hearing loss (family history, premature birth, intensive care stay, meningitis, congenital infection such as cytomegalovirus, craniofacial anomalies, and exposure to ototoxic drugs).
- Children and adults who cannot do behavioural hearing tests (young children, learning disability, non-cooperative patients).
- Detection of auditory neuropathy spectrum disorder, where emissions are present but the auditory brainstem response is abnormal or absent.
- Monitoring of hearing in patients on ototoxic drugs (aminoglycosides, cisplatin, loop diuretics), including high-frequency DPOAE, which change before pure tone hearing.
- Monitoring of noise exposure effects in workers and musicians.
- Distinguishing cochlear from retrocochlear loss, and suspected non-organic hearing loss (emissions present with reported severe loss).
- Part of the test battery in tinnitus, sudden hearing loss, and in research.
- Not a test of hearing in the full sense: it does not measure the hearing threshold, or the function of the nerve and brain.
Contraindications and precautions.
- No contraindications.
- Wax, vernix (in newborns in the first hours after birth), debris, or fluid in the ear canal, and middle ear fluid or infection, cause a refer result without hearing loss, so ears are examined and the test is repeated.
- The probe is placed carefully in infants, with disposable tips, and the test is not done in painful or infected ears, and discharging ears.
- Noise in the room, and a restless, crying, feeding or sucking baby, increase background noise and cause a false refer result.
- Recent loud noise exposure reduces emissions temporarily.
- Patients with very small or collapsed canals may need a different tip.
- Equipment is calibrated, and probes are checked daily, since blocked probes give false refer results.
Preparation and technique.
Otoscopy is done, and any visible wax and debris are removed gently. In a baby, the test is done when the child is asleep or quiet, usually after the first feed, within a quiet room, in the first 1 to 2 days of life (and after 24 hours, since vernix and fluid clear). The probe tip is placed in the ear canal with a good seal, and the stimulus level and noise level are checked on the screen.
The device presents clicks (TEOAE) or tone pairs (DPOAE), and records and averages the responses. A pass requires a reproducible response and a signal-to-noise ratio of about 6 dB or more at several frequencies (for example, 3 of 4 or 5 frequencies for DPOAE, and reproducibility of over 70% for TEOAE), depending on the device settings.
If the result is refer, the test is repeated, and the ear is rechecked.