Back

Auto-PEEP Syndrome

Auto-PEEP (intrinsic positive end-expiratory pressure) is end-expiratory alveolar pressure above the pressure set on the ventilator, because exhalation is incomplete before the next breath begins. Gas accumulates (breath stacking), and lung volume at the end of expiration rises. This article concentrates on the ventilated patient.

Causes

  • Short expiratory time: high respiratory rate, large tidal volume, long inspiratory time, high minute ventilation
  • Expiratory airflow obstruction: asthma, chronic obstructive pulmonary disease, bronchospasm, mucus plugs, small endotracheal tube, kinked tube and high-resistance expiratory filters or heat-moisture exchangers
  • High lung compliance and expiratory valve problems

Consequences

  • Raised intrathoracic pressure that reduces venous return and cardiac output, which causes hypotension and can cause pulseless electrical activity (frequent after intubation of a patient with severe asthma)
  • Increased work of breathing, with inspiratory efforts that fail to trigger the ventilator
  • Barotrauma and volutrauma
  • False elevation of central venous and pulmonary artery occlusion pressures
  • Misleading plateau pressure and compliance measurements

Severe cases are fatal within minutes. Complications: cardiovascular collapse, pneumothorax, ineffective triggering with patient-ventilator asynchrony, and hypercapnia.

Related

Clinical toolsCalculators

Latest content

Atlas’ Videos

On the evidence

Doses are traceable.

Management sections cite the guideline they came from. Atlas supports clinical judgement rather than replacing it; verify against current national guidance and the patient in front of you.

Learn it.
Know it.
Own it.