Electroretinography

Electroretinography records the electrical response of the retina to light. It measures the function of the photoreceptors (rods and cones) and the cells that connect to them, such as the bipolar cells, and it detects disease of the retina before changes are visible on examination. It is an objective test that does not depend on the patient's response.

Light flashes of set intensity and colour are presented to the eye, and the response is recorded with electrodes on the cornea or lower lid (contact lens, thread or foil electrodes) and skin electrodes near the eye. The test has several parts.

The full field electroretinogram measures the response of the whole retina, and is done after 20 to 30 minutes of dark adaptation for rod and combined responses, and after light adaptation for cone responses and flicker. The pattern electroretinogram measures the function of the macula and retinal ganglion cells, using a reversing checkerboard.

The multifocal electroretinogram maps the responses from about 100 small areas of the central retina. The electro oculogram measures the function of the retinal pigment epithelium. Visual evoked potentials may be added to test the visual system as far as the brain.

The test takes 30 to 90 minutes. The main waves are the a-wave (from photoreceptors), the b-wave (from bipolar and supporting glial cells), and oscillatory potentials (from inner retinal cells). Full field electroretinography is abnormal when more than about 20 to 30% of the retina is damaged, so a focal macular disease may be missed. The test confirms the diagnosis in inherited retinal dystrophies in most patients, and separates rod cone from cone rod patterns.

Indications.

  • Suspected inherited retinal dystrophy: retinitis pigmentosa, cone rod dystrophy, Leber congenital amaurosis, congenital stationary night blindness, achromatopsia, and Stargardt disease.
  • Unexplained visual loss with a normal appearing fundus, night blindness, and poor vision in infants and young children.
  • Suspected toxic retinopathy (hydroxychloroquine, vigabatrin, other drugs) and retinal damage after poisoning.
  • Retinal inflammation and degeneration with unclear cause, including autoimmune retinopathy and paraneoplastic retinopathy.
  • Vascular disease: central retinal vein and artery occlusion (to assess ischaemia), and diabetic retinopathy in research.
  • Before gene therapy, and to document baseline function.
  • Media opacity (dense cataract, vitreous haemorrhage) when the retina cannot be seen, to assess retinal function before surgery.
  • Medico legal documentation of retinal function, and suspected non-organic visual loss.

Not indicated for diseases that are clear on examination and imaging, and not sensitive to small, focal lesions.

Contraindications and precautions.

  • Corneal abrasion, infection, or a very fragile cornea: the corneal electrode is avoided or a skin electrode is used.
  • Open globe injury or recent eye surgery: delayed until healed.
  • Allergy to topical anaesthetic or electrode gel or adhesive.
  • Photosensitive epilepsy: caution with flashes.
  • Pupil dilation is needed, with the cautions for narrow angles.
  • Infants and young children need sedation or anaesthesia, which affects the response, so the same anaesthetic and a controlled setup are used.
  • Patients taking drugs that affect the response (some anticonvulsants, anaesthetic agents) need notes made.

Preparation and technique.

The pupils are dilated fully, and the patient is dark adapted for 20 to 30 minutes. Corneal anaesthetic drops are given, and electrodes are placed: an active electrode on the cornea, a reference near the outer corner of the eye, and a ground on the forehead or earlobe. The patient sits in front of a stimulator bowl and looks at a fixation point.

Flashes of increasing intensity are given to record the rod response, then the maximal combined response. The patient is then light adapted for 10 minutes, and cone responses and 30 Hz flicker are recorded. Several responses are averaged, and artefacts from blinks and movements are rejected. The recording is repeated to confirm reproducibility.

Results are compared with the laboratory's age matched normal values.