Ocular Ultrasound

Ocular ultrasound uses high frequency sound waves to produce images of the eye and orbit. It is the main way of examining the back of the eye when the view is blocked by a dense cataract, corneal opacity, vitreous haemorrhage, or a small pupil, and it also measures the length and structure of the eye and the tissues of the orbit. It is quick, painless, and needs no contrast.

Two modes are used. A-scan (amplitude) is a one dimensional trace, with spikes whose height shows the strength of the echo, and whose position shows the depth. It measures axial length for lens implant calculation, and shows the internal reflectivity of a lesion.

B-scan (brightness) gives a two dimensional, cross-sectional picture, with probes of about 10 MHz that image the retina, vitreous, choroid and orbit. Ultrasound biomicroscopy uses 35 to 50 MHz probes, which give a higher resolution but a depth of only 4 to 5 mm, and images the cornea, angle, iris, ciliary body and anterior lens.

Doppler ultrasound shows blood flow in the retinal and orbital vessels.

The patient lies back, and the probe is placed over the closed upper lid with a coupling gel, or on the cornea with an immersion technique for biomicroscopy. The test takes 5 to 15 minutes. The normal axial length is about 23 to 24 mm in adults, and an error of 1 mm gives a lens power error of about 2.5 to 3 dioptres.

The test detects retinal detachment with an accuracy of over 95% in experienced hands, and is the main tool for the diagnosis and follow up of intraocular tumours when the fundus cannot be seen.

Indications.

  • Opaque media (dense cataract, vitreous haemorrhage, corneal scar, hyphaema, small or occluded pupil): to find retinal detachment, tumour, or other disease before surgery.
  • Suspected retinal or choroidal detachment, and vitreous haemorrhage, with the cause unknown.
  • Intraocular tumours: choroidal melanoma, haemangioma, metastasis, and retinoblastoma (with calcification), and their measurement over time.
  • Eye trauma: foreign bodies, vitreous haemorrhage, lens dislocation, and retinal detachment, in a closed globe.
  • Orbital disease: thyroid eye disease, orbital mass, cellulitis, abscess, and optic nerve sheath enlargement.
  • Axial length and anterior chamber depth measurement for lens implant calculation, and for follow up of myopia.
  • Optic disc drusen and papilloedema, and raised intracranial pressure by measuring the optic nerve sheath diameter.
  • Anterior segment disease with ultrasound biomicroscopy: angle closure and plateau iris, ciliary body tumours, cysts, and implant position.
  • Unexplained proptosis and glaucoma with an unclear cause.

Not suitable alone when the eye can be seen well and optical coherence tomography is available, or when an open globe injury is suspected.

Contraindications and precautions.

  • Suspected open globe injury (a ruptured eye): no pressure on the eye, and the test is delayed or done with the lightest touch and a thick layer of gel. CT is used first.
  • Recent eye surgery or wounds: care with the probe and sterile gel.
  • Active infection of the lids: the probe is cleaned and covered.
  • Immersion biomicroscopy needs topical anaesthesia and a clean cup.
  • Allergy to gel (rare).
  • Patients who cannot lie flat or keep still: scanning in a seated position, or sedation in children.
  • The B-scan probe is cleaned between patients with disinfectant to prevent spread of infection.

Preparation and technique.

The patient lies supine or sits, with the eyes closed, and coupling gel is applied to the lid. The B-scan probe is placed over the closed lid, and the examiner scans in horizontal, vertical and oblique planes, with the patient looking in different directions (kinetic scanning) to observe how structures move.

The gain is changed to detect weak echoes (vitreous opacities, haemorrhage) and strong echoes (membranes, calcified lesions). Both eyes are compared. For A-scan biometry, the probe touches the anaesthetised cornea, or an immersion shell is used, and 5 or more measurements are taken. For optic nerve sheath measurement, the diameter is measured 3 mm behind the globe.

The images and measurements are saved.