Fluoroscopy (General)
Fluoroscopy is the real time radiographic imaging of moving structures. It uses a continuous or pulsed x-ray beam and an image detector, usually a flat panel digital detector, with images displayed on monitors.
It is the basis of contrast studies (swallows, enemas, urography, cystography, arthrography, myelography and fistulography).
It is also the basis of procedures under image control (needle placement, drain and stent insertion, angiography, vascular and non-vascular interventions, pain injections, line placement, fracture reduction, orthopaedic surgery, endoscopic retrograde cholangiopancreatography, pacemaker and cardiac catheter lab work).
It supports functional assessment (diaphragm movement, joint motion, swallowing, and gastric emptying).
Images are acquired as continuous fluoroscopy (low dose rate), as pulsed fluoroscopy (3 to 15 pulses each second, lowering dose), as digital spot images and cine (higher dose), and as digital subtraction images.
The last image is held on screen after exposure ends (last image hold) so that the operator can review it without more radiation.
The dose rate in ordinary fluoroscopy is about 10 to 30 mGy each minute at the patient entrance surface. It may reach 100 mGy each minute or higher in large patients or in a high dose mode. Cine acquisition is 10 to 60 times higher than fluoroscopy.
Fluoroscopy is one of the largest contributors to medical radiation exposure for staff and patients, as procedure times can be long. Effective doses range from 1 to 10 mSv for diagnostic studies to 10 to 50 mSv or more for complex interventional procedures.
Staff doses are due to scatter from the patient, so the operator is exposed to the highest dose.
Deterministic effects (skin burns, hair loss, cataracts) occur at these skin doses:
- about 2 Gy: early erythema
- 3 to 5 Gy: epilation and moist desquamation
- over 10 Gy: necrosis
Cataract risk in staff is related to the cumulative lens dose. The principles are justification, optimisation (dose as low as reasonably achievable), and dose limits for staff.
Indications. Any examination or procedure where real time or dynamic imaging is needed:
- dynamic contrast studies of the gastrointestinal, urinary, genital and biliary tracts
- joint and spinal injections and arthrography
- vascular access and angiography
- drainage and biopsy under image control
- line, catheter, tube and drain placement and exchange
- orthopaedic fracture reduction and fixation in theatre
- swallow assessment
- assessment of diaphragm movement (sniff test) and tracheal motion
- localisation of foreign bodies
- pacing wire and cardiac device work
- endoscopic retrograde cholangiopancreatography
- ureteroscopy and stone treatment
- check of tubes and prostheses
Not when ultrasound, MR or a still radiograph answer the question without ionising radiation.
Contraindications and precautions.
- Pregnancy: avoid or use ultrasound or MR. When unavoidable, keep the beam away from the uterus and shorten the exposure.
- Children: lower dose settings, pulsed fluoroscopy at low rates, tight collimation, and no anti-scatter grid for small children.
- Previous high skin dose: avoid the same entry field.
- Allergy to the contrast agent where contrast is used.
- Women of reproductive age: a pregnancy test or a check of the menstrual dates as local policy requires.
- Unstable patients: monitoring and resuscitation equipment needed.
Staff protection:
- lead aprons (0.25 to 0.5 mm lead equivalent)
- thyroid collars
- lead glasses (reduce lens dose by a factor of about 5)
- ceiling suspended shields and table side curtains
- dosimeters at collar level and under the apron
- training
Preparation and technique.
- The patient is positioned on the table.
- The field is collimated tightly to the region of interest.
- The detector is placed as near as possible to the patient and the x-ray tube as far as possible.
- The lowest dose rate and pulse rate that give an adequate image are chosen, and magnification modes are avoided.
The beam angle is varied during long procedures to spread the skin dose over different areas. Fluoroscopy time and dose area product are recorded, and an alert is set at cumulative air kerma of 3 to 5 Gy.
Operators stay as far from the beam as practical, stand on the side of the detector (low scatter) rather than the tube side, and wear protection.