Point-of-Care Cardiac Markers

Point-of-care (POC) cardiac marker testing measures cardiac troponin I or T (and sometimes creatine kinase MB, myoglobin, B-type natriuretic peptide, N-terminal pro-BNP and D-dimer) in whole blood, plasma or serum at the bedside or in the emergency department, with results in 10 to 20 minutes.

Conventional POC troponin assays are less sensitive than central laboratory high sensitivity troponin (limit of detection 10 to 50 ng/L, compared with 1 to 5 ng/L), and need serial sampling at presentation, 3 hours and sometimes 6 hours.

High sensitivity POC troponin systems (limit of detection 1 to 5 ng/L, 99th percentile 10 to 30 ng/L) now allow rapid rule out at 0 and 1 to 3 hours with negative predictive value over 99%, where validated.

Chest pain accounts for 5 to 10% of emergency visits, and acute myocardial infarction is diagnosed in 10 to 20% of those presenting with suspected acute coronary syndrome.

Troponin is the biomarker of choice for myocardial injury, and the diagnosis of myocardial infarction requires a rise and fall in troponin with at least one value over the 99th percentile upper reference limit and symptoms, ECG changes, imaging evidence or thrombus.

B-type natriuretic peptide supports diagnosis and prognosis in heart failure (acute breathlessness), and D-dimer helps exclude pulmonary embolism and aortic dissection in low risk patients.

Indications. Suspected acute coronary syndrome (chest pain, anginal equivalents, breathlessness, syncope, arrhythmia). Breathlessness with suspected heart failure (BNP or NT-proBNP). Suspected pulmonary embolism in low probability patients (D-dimer). Pre-hospital and rural settings where a laboratory is distant. Triage of undifferentiated emergency presentations.

Post-operative, post-procedure or chemotherapy cardiac monitoring. Myocarditis, pericarditis and cardiotoxicity assessment. Not to be used alone, without ECG and clinical assessment.

Specimen. Whole blood from a finger prick or venous sample (EDTA or lithium heparin, 0.15 to 0.25 mL, cartridge dependent), plasma or serum. Time of symptom onset and sample times are recorded, since troponin rises 2 to 3 hours after injury, peaks at 12 to 24 hours and remains elevated for 7 to 14 days. Strict adherence to the manufacturer's time and temperature requirements.

Repeat samples at 1 to 3 hours (high sensitivity assays) or 3 to 6 hours (conventional) after presentation, or later if symptoms continue.