Point-of-Care Haemoglobin

Point-of-care (POC) haemoglobin testing estimates haemoglobin concentration in a drop of capillary or venous blood within 1 minute, using portable photometers (azide methaemoglobin reaction on a microcuvette, read at 570 and 880 nm), digital haematology analysers, blood gas analysers (calculated from haematocrit) and non-invasive spectroscopic devices (pulse co-oximetry).

Photometric devices have accuracy within 0.5 to 1.0 g/dL of laboratory values, are the usual field method for anaemia surveys and donor screening, and the colour scale or Sahli methods are no longer accepted. Non-invasive devices are convenient but less precise (limits of agreement up to plus or minus 2 g/dL), and are used for screening, not decisions on transfusion.

Anaemia affects about 1.9 billion people (nearly a quarter of the world), including 40% of children under 5 years and 30 to 40% of pregnant women in low income countries, and is the leading cause of years lived with disability among women.

Iron deficiency is the commonest cause, followed by infection (malaria, hookworm, HIV, tuberculosis), haemoglobinopathies, vitamin B12 and folate deficiency, chronic disease and blood loss. Rapid haemoglobin supports transfusion decisions in haemorrhage, antenatal care, malaria, sickle cell crisis, surgery and neonatal practice.

Indications. Pallor, fatigue, breathlessness, tachycardia, heart failure or syncope. Antenatal booking and each trimester. Postpartum haemorrhage and obstetric emergencies. Severe malaria and malnutrition. Trauma, gastrointestinal bleeding and surgery. Blood donor screening. Child health clinics and school surveys.

Monitoring in chronic kidney disease, cancer treatment, sickle cell disease, and iron or erythropoietin therapy. Pre-operative assessment in remote settings. Not for diagnosis of the cause of anaemia.

Specimen. Capillary blood from the side of a fingertip, from the ring or middle finger (the second drop after wiping away the first, without squeezing, which dilutes with tissue fluid), heel in infants, or venous or arterial whole blood in EDTA or heparin.

Fill the microcuvette in one continuous motion without air bubbles, wipe the outside, and read within 40 seconds to 1 minute (the instrument stated time). Cuvettes must be stored sealed and dry, because moisture alters the reagent. Record altitude, smoking, and pregnancy for interpretation.