D-Dimer
D-dimer is a degradation product released when plasmin breaks down cross-linked fibrin, and its presence shows that coagulation and fibrinolysis have both been activated. It is measured on citrated plasma by latex enhanced immunoturbidimetric or enzyme immunoassay methods.
Its main use is as a rule out test. A negative result with a low or intermediate pre-test probability makes acute venous thromboembolism (deep vein thrombosis and pulmonary embolism) very unlikely, with a negative predictive value over 95%.
A positive result is not diagnostic, because many conditions raise it.
Units. Results are reported as fibrinogen equivalent units (FEU) or D-dimer units (DDU). Numerically, FEU values are about double DDU values. The cut off must match the unit, and a mismatch is a source of serious error.
Specificity falls with age, with only a small minority of people over 80 years having a result under the conventional cut off. It is also low in hospital inpatients, cancer, infection, pregnancy, trauma and after surgery.
Specimen. Citrated plasma. Samples are stable for about 4 to 8 hours at room temperature, and longer in some assays. Haemolysis, lipaemia and icterus can interfere.